Top Grade Baby Infant Incubator Equipment
1Air mode and baby mode controlled by micro-computer;
2Humidity control system and Oxygen concentration control system;
3Set temperature, air temperature, baby temperature, humidity and oxygen are displayed separately;
>37ºC temperature set function;
5Failure alarm indication;
6Embedded integrated sensor box;
7integrated water reservoir in the shape of drawer;
8Double wall hood, automatic air circulation device;
9Inclination of bassinet is adjustable;
10RS-232 connector.
11Weighing system and Monitor shelf are optional.
Standard configuration
The hood, the base, the mobile cabinet, the controller, I. V pole, shelf; Oxygen concentration control system.
Optional configurations:
1. Fixed Neonate bilirubin phototherapy equipment (with blue flourencence lamps)
2. Fixed Neonate bilirubin phototherapy equipment (with LED blue lights)
3. Vertical Height Adjustment Stand (VHA Stand) Weighing System, DC motor, disposable skin temperature sensor.
Power supply AC220V-230V/50Hz or AC110-120V/50-60Hz
Max Power Output 1000VA
Mode of control Air mode and baby mode controlled by micro-computer
Air temperature control range 25C-37C(override mode 37-39C)
Baby temperature control range 34-37C(override mode 37-38C)
Accuracy of skin temperature sensor: ±0.3C
Temperature variability ≤0.5C
Temperature uniformity ≤0.8C(mattress is on the horizontal position)
Inclination of the bassinet ±12°
Noise inside hood ≤55dB(A) (Environment noise≤45dB(A))
Optional: with DC motor. Noise inside hood ≤45dB(A) (Environment noise≤35dB(A))
Failure alarm over temperature alarm, deviation alarm, sensor failure alarm, fan motor alarm, power failure alarm, sensor box position alarm and system alarm,low water alarm and tank position alarm
Humidity display range 0%RH~99%RH
Humidity control range 0%RH~90%RH
Accuracy of humidity ±10%RH
Indication range of oxygen concentration 0%~99%
Control range of oxygen concentration 20%~60%
Accuracy of oxygen concentration ±3%